Manual external carotid artery manipulation for failed device advance through the stent in carotid artery stent placement. a technical note
Main Article Content
Abstract
We report herein the usefulness of a manual external carotid artery manipulation for failed advancement of
devices through the stent in carotid artery stent (CAS) placement with open cell type stent and filter type
distal protection device. In all, 22 consecutive patients underwent CAS with filter type protection device
between April 2008 and December 2009 in our institution, and failed advancement of the devices through
the stent occurred in 4 patients (18%). For all the 4 patients, the devices could be navigated normally
through the stent under the manual external carotid artery manipulation. In cases with failed device advance
in CAS, this maneuver would be one of the methods to resolve this.