Flow-independent dynamics in aneurysm (FIDA): pressure measurements following partial and complete flow impairment in experimental aneurysm model

Description

Background: There have been growing concerns regarding delayed aneurysm rupture subsequent to the flow-diverting stent deployment. Therefore, more investigations are needed regarding hemodynamic changes secondary to flow-diverting stent deployment. Objective: To study intra-aneurysmal and perianeurysmal pressures after partial and complete flow impairment into the aneurysm. Methods—A silicone model of an 8-mm-sized aneurysm (neck diameter: 5 mm, vessel size: 4 mm) was used. The aneurysm wall was encapsulated and sealed within a 5 ml syringe filled with saline and a pressure sensor guide wire (ComboWire, Volcano Corp.) to detect pressure changes in the perivascular compartment (outer aneurysm wall). A second pressure sensor guide wire was advanced inside the aneurysm sac. Both pressure sensors were continuously measuring pressure inside and outside the aneurysm under pulsatile flow under the following conditions: 1) baseline (reference); 2) a 16 mm by 3.75 mm flow-diverting stent (ev3/Covidien Vascular, Mansfield, MA) deployed in front of the aneurysm; 3) two flow-diverting stents (16 mm by 3.5 mm) were deployed; and 4) a covered stent (4 mm by 16 mm VeriFlex coronary artery stent covered with rubber sheet) was deployed. Results: Mean (±SD) baseline pressures inside and outside the aneurysm were 53.9 (±2.4) mmHg (range 120–40 mmHg) and 15.4 (±0.7) mmHg (range 40–8mmHg), respectively. There was no change in pressure inside and outside the aneurysm after deploying the first and second flow-diverting stents (partial flow impairment) and it remained at 53.9 (±2.7) mmHg and 14.9 (±1) mmHg for the pressure inside and outside the aneurysm, respectively. The pressure recording from outside the aneurysm dropped from 15.4 (±0.7) mmHg to 0.3 (±0.7) mmHg after deploying the covered stent (complete flow impairment). There was no change in pressure inside the aneurysm after deploying the covered stent. Mean (±SD) pressure within the aneurysm was 55.1 (±1.7) mmHg and it remained 54.7 (±1.7) mmHg after covered stent deployment.

Files

jvin7-3-11.pdf

Files (840.4 kB)

Name Size Download all
md5:69a81109f26c8ad5178cff8a80ebf0a6
840.4 kB Preview Download