Published November 12, 2023 | Version v1
Publication Open

Initial Clinical Experience with a New Self-Expanding Nitinol Microstent for the Treatment of Wide-Neck Intracranial Cerebral Aneurysms: the Acandis Acclino Stent

Description

Background and Purpose—The Acclino is a laser-cut closed-cell microstent composed of nitinol. It was developed for stent-assisted coiling of wide-neck intracranial aneurysms. The key feature of the stent is its deployability via low-profile microcatheters with an inner diameter of 0.0165 inch, which are also suited for coil deployment. The objective of this study was to evaluate the safety and feasibility as well as the immediate and mid-term results of this new device. Materials and Methods—Our database was screened for all Acclino-based stent-assisted intracranial coil embolizations since its introduction to the European market in June 2012. Case files and imaging data were retrospectively analyzed for angiographical and clinical outcome parameters, including immediate and mid-term modified Raymond-Roy aneurysm occlusion classification (RROC) rates and procedural complications. Results—Fourteen patients comprising 14 aneurysms (9 unruptured and 5 ruptured) were treated with the Acclino. All except for a dissecting one were wide-neck saccular aneurysms. Immediate complete occlusion (RROC1) was observed in 8/14 cases (57%), a residual neck (RROC2) in 4/14 (29%), and a persistent filling of the dome (RROC 3) in 1/14 cases (7%). An in-stent thrombus formation in one case (7%) was medically resolved without neurological deficit. Follow-up was available in 9/14 cases (64%) after a mean of 137 days (SD ± 50). All followed cases depicted a complete occlusion (RROC1)

Files

jvin-8-3-1.pdf

Files (1.0 MB)

Name Size Download all
md5:685faf732fee4b7fdeffedb50c78a6a4
1.0 MB Preview Download